FDA Classifies Intersurgical Breathing Tube Recall Most Serious
The Food and Drug Administration has classified Intersurgical's removal of its 22mm Smoothbore breathing tubes and systems as a Class I recall, its most serious type, after company testing found some connectors were bonded with too little solvent and can detach during use. The company reported no serious injuries or deaths as of November 3, 2026.
Edited and fact-checked by Stephen Sprague
Hospitals must pull twelve Intersurgical Smoothbore product codes from service and match every unit against the FDA's lot list, because a connector bonded with too little solvent can detach mid-ventilation.
A breathing circuit in an intensive care unit does one job. It carries air and oxygen from a ventilator to a patient and lets exhaled air flow back out, and Intersurgical's 22mm Smoothbore tubes and systems are built for exactly that work. The soft connectors at the ends of the tubes are the points where the circuit seals, and the Food and Drug Administration says some of those connectors can come off the tube while a patient is being ventilated.
The agency classified the company's removal of the devices as a Class I recall on November 20, 2026, and updated the recall communication on November 25, after issuing an Early Alert about the issue on November 10. Class I is the FDA's most serious recall category, reserved for situations where continued use of a device may cause serious injury or death. The failure mode here is mechanical, a bonded joint that can separate, and it sits in equipment that a patient can't breathe without.
Intersurgical told the FDA that a solvent called tetrahydrofuran is used during manufacturing to bond the soft connectors onto the Smoothbore tube. Reports from markets outside the United States described soft connectors disconnecting from the tube, and the company's investigation and testing of its internal stock found that some products were manufactured with too little solvent to ensure a secure connection. Because the defect sits inside the joint, a tube can look intact on inspection and still pull apart under the movement and handling of ordinary clinical use.
A connector that disconnects during use can cut off ventilation or leave it compromised, and the patient desaturates until a clinician restarts ventilation manually. Depending on the individual patient, the specific situation and the time needed to reinitiate ventilation, the FDA says that loss of ventilation could lead to organ failure, brain damage or death caused by hypoxia. As of November 3, 2026, Intersurgical had reported no serious injuries or deaths associated with the issue, so the classification rests on the tested failure and the disconnection reports from outside the United States.
Twelve product codes are covered across the Smoothbore range. The list runs from a plain 22mm tube with soft end connectors to single and dual heated wire breathing systems with water traps, systems with monitoring lines and exhalation valves, and bilevel systems with swivel carbon dioxide leak ports, one of them fitted with a Clear-Guard Midi filter. Lengths are at or above 1.6 meters or 1.8 meters depending on the model, and the FDA lists the affected lot numbers for every code, from a single lot on some models to more than thirty lots on the filtered bilevel system.
Intersurgical sent its customer letter on November 3, 2026, and the instructions are direct. Facilities must immediately discontinue use, quarantine any stock with the affected product and lot numbers, and return the product to Intersurgical, and the company also told customers to share the notice with all users inside their facility and to contact their own customers immediately if they distributed the products further. United States customers with adverse reactions, quality problems or questions can contact Intersurgical Customer Service at [email protected] or 800-828-9633, and health professionals and consumers can report problems to MedWatch, the FDA's safety reporting program.
What the FDA page doesn't give hospitals is a unit count. It states no total for the devices shipped into United States facilities and no distribution map for the affected lots, so each facility has to match its own shelves, storage rooms and fitted ventilators against the published lot list, code by code. A breathing system already connected to a patient carries the same instruction as one still sealed in its packaging, and both come out of service until the lot check says otherwise.