$2 / month

Comment and reply across TML.

Start or manage membership

Most pulsed

Latest from other desks

tml.the modern leap
Health / Medical Research

FDA Flags Portex Breathing Tubes After One Death

The Food and Drug Administration issued an early alert on November 25 saying ICU Medical is removing certain Portex endotracheal tubes because a resin problem can leave the tube material softer than expected. The small uncuffed tubes are made for newborns and young infants, and the company had reported one death associated with the issue as of November 16.

TML News DeskNov 26, 12:36 AM ET · 2 min read

Edited and fact-checked by Stephen Sprague

An endotracheal tube in an archive photo. The recalled Portex tubes are the small uncuffed sizes made for newborns and young infants. Photo: bigomar2, CC BY-SA 3.0, via Wikimedia Commons.
Summary

Any hospital still holding the listed Portex lots should pull them from stock now, because the FDA says a soft tube can kink during a newborn's intubation and cut off ventilation.

Tap a passage. React to the part that matters.

A breathing tube that's softer than intended can kink while a clinician guides it into a newborn's windpipe, cutting off airflow during an emergency. On November 25, 2026, the Food and Drug Administration issued an early alert warning hospitals that ICU Medical is removing specific lots of its infant Portex endotracheal tubes after discovering a manufacturing defect in the tubing resin.

According to the agency, affected lots contain inconsistent levels of plasticizer, leaving the flexible plastic softer than specifications allow. In tiny uncuffed tubes sized for newborns and young infants, that extra softness can increase resistance to ventilation, make intubation harder, and cause the tube to fold or kink during placement. If a care team doesn't immediately spot the obstruction, an infant can become inadequately ventilated and require an emergency tube replacement, a failure the FDA warns can contribute to severe injury or death.

As of November 16, ICU Medical had reported one death associated with the problem. Under federal device reporting rules, an associated death means a fatality was reported alongside the product issue, and the FDA's alert doesn't establish that the tube caused the death.

The removal covers five uncuffed Portex catalog models in 2.5-, 3.0-, and 3.5-millimeter diameters, including three Murphy eye oral and nasal tubes (items 100/127/025, 100/127/030, and 100/127/035) and two preformed Polar Clear south oral tubes shaped for head, neck, and mouth surgery (items 100/134/030 and 100/134/035), with specific lot numbers posted on the FDA's alert page.

ICU Medical notified customers by letter on November 20, instructing hospitals to check every storage location, immediately stop using the listed lots, and destroy them under institutional protocols. Facilities that can't destroy the stock right away must quarantine the tubes and post the recall notice directly on the bins so no clinician grabs one during an emergency, while also forwarding the warning to any secondary distributors.

Regulators have flagged the issue as potentially high-risk while continuing their review and haven't yet assigned a formal recall class. Clinicians and hospitals can report adverse events or tube failures to ICU Medical's complaint team at 866-216-8806 or through the FDA's MedWatch reporting program.

The conversation

A little more perspective.

Reading is free. Verified members can comment and reply for $2 a month.

The conversation.

Loading comments…

Keep your curiosity going.

Some links here are affiliate links. If you buy through one, we might make a commission — it never changes what you pay. As an Amazon Associate I earn from qualifying purchases.