FDA Flags Tandem Mobi Recall As Most Serious Type
The FDA published its Tandem Mobi insulin pump recall notice on November 25, 2026, and identified it as the most serious type, after Tandem Diabetes reported 1,248 serious injuries tied to false alarms that can suspend insulin delivery. Some pumps are replaced by serial number, and the rest get updated instructions.
Edited and fact-checked by Stephen Sprague
The FDA calls this its most serious recall type. Check the serial number on the back of the pump: listed pumps get replaced through Tandem at 877-801-6901, and every other Mobi user follows the updated alarm instructions and keeps a backup insulin method ready.
The alarm sounds like any other occlusion warning, and the pump stops delivering insulin until its owner clears it.
On the Tandem Mobi, that alarm can be false, because Tandem Diabetes says certain pumps generate false-positive occlusion alarms when no blockage exists, and delivery stays suspended until the user acknowledges the alarm and resumes therapy.
The Food and Drug Administration published its recall page on November 25, 2026, and identified the recall as the most serious type, the category for devices that may cause serious injury or death. As of October 28, Tandem had reported 1,248 serious injuries associated with the issue and no deaths, and the recall affects about 80,900 pumps, almost 78,000 of them in the United States.
A suspended pump delivers no insulin, and repeated false alarms that go unanswered can leave blood glucose climbing, because interruptions of insulin delivery raise the risk of elevated blood glucose and severe hyperglycemia may require medical intervention or hospitalization.
The recall splits owners by serial number, since pumps on the FDA's replacement list come out of use and their owners submit a return and replacement request through the link in Tandem's customer letter or by calling Tandem Customer Technical Support at 877-801-6901. Every other Mobi pump with interoperable technology stays in use under updated instructions that Tandem sent in a separate letter.
Both letters went out on September 23, 2026, and both tell users to answer every occlusion alarm, inspect the cartridge, tubing and infusion site for damage or blockage, and keep the Mobi phone app's push notifications enabled so an alarm also reaches the phone. The FDA also tells users to check blood sugar regularly and to keep a backup method of insulin delivery ready, as directed by their healthcare provider.
The recall covers all Tandem Mobi insulin pumps with interoperable technology, the pumps that deliver insulin under the skin and pair with compatible continuous glucose monitors and automated dosing systems to adjust delivery. Healthcare providers, distributors and pharmacies received letters as well, and problems with the pump can be reported to FDA MedWatch, the agency's adverse event reporting program.
The FDA page doesn't describe the 1,248 injuries case by case, giving the count, the date, and the mechanism instead, and the agency says it will update the page as significant new information becomes available. The replacement decision is also narrower than the headline suggests, because only the serial numbers on Tandem's list come back, while a pump that isn't listed stays in service under the updated instructions.
For a person who sleeps wearing the pump, the practical change is small and nightly: notifications on, backup insulin within reach, and a serial number checked against the replacement list.