FDA Posts Luna BPAP Correction Over Wrong Pressure Data
The FDA has posted a correction for 7,533 Luna G3 BPAP devices that record wrong pressure statistics when a feature called Targeted VT is on. The delivered pressure stays in range, the written history doesn't, and a firmware update on SD cards is the fix.
Edited and fact-checked by Stephen Sprague
The therapy was right and the record was wrong. For a device a clinician tunes from its own history, the record is half the treatment.
The machine delivers the right pressure and writes down the wrong one. The Food and Drug Administration posted a correction update on November 19, 2026, for the Luna G3 BPAP System, model LG3800-30VT, a bilevel airway pressure device made by React Health, doing business as 3B Medical. The update covers 7,533 devices distributed in the United States and Puerto Rico, and the defect sits in the therapy record the device keeps, not in the air it pushes.
With a feature called Targeted VT switched on, the device doesn't record the actual pressure statistics it delivered, either on its internal SD card or in the cloud system, React Health Connect, that clinicians use to review therapy. The pressure itself stays within the minimum and maximum the clinician set. The history doesn't. A clinician reading that history sees numbers that don't fluctuate the way the real therapy did, and adjustments made from a wrong record can send the therapy in the wrong direction or delay a correction the patient needs.
React Health first told customers about the problem on October 8, 2026, and the company says it has received 15 complaints with two serious adverse reports tied to the issue. Those counts are the company's own, reported in the FDA posting, and the agency notes it doesn't endorse either the product or the company. The fix is a firmware update, version G3-2.00.80, arriving on SD cards through the equipment providers who service the machines. The update doesn't repair the historical record. Targeted VT data already stored stays wrong, and the company's original instructions direct users to erase the stored history after updating.
The instructions split by who's reading them. Clinicians are told to stop using Targeted VT on affected devices right away. Patients get the opposite instruction about their therapy itself. "Do not independently change your prescribed therapy settings or discontinue respiratory therapy without guidance from your health care provider," the company says, and patients should keep using the device as prescribed until their provider advises otherwise. React Health's customer support line, 863-226-6285, staffed Monday through Friday from 8 a.m. to 5 p.m. Eastern, is the contact the posting gives for questions.
A correction rather than a removal means the devices stay in bedrooms while the firmware catches up, and the gap in the record stays with them. Anyone whose clinician adjusted therapy from Luna G3 data this fall has a concrete question for their next appointment, because the numbers behind those adjustments may not describe the pressure the machine actually delivered.
A bilevel device like the Luna G3 delivers two pressure levels, a higher one as the patient breathes in and a lower one as they breathe out, and clinicians tune those levels over time from the machine's own usage record. That record is what this correction is about. When the stored statistics don't match the therapy, every later adjustment inherits the error, and the erase instruction means the faulty history won't sit in the file as a warning. It'll be gone, and the corrected firmware starts the record over from the update date.