FDA Warns Guidewire Coating Could Enter Patient Bloodstream
Heraeus Medevio told customers on November 4-5 to stop using the affected lots and quarantine them. The FDA calls the issue potentially high-risk and says its review is still open.
Stop use. Quarantine the lots.
The FDA says certain Heraeus Medevio vascular guidewires shouldn't be used because their coating may clump, detach and enter a patient's bloodstream.
The agency's Center for Devices and Radiological Health issued the Early Alert on November 20, 2026, after Heraeus Medical Components, doing business as Heraeus Medevio, sent a voluntary recall letter to affected customers on November 4-5. The affected products are Navilyst vascular guidewires (models H965457461 and H965457471), Spectrum vascular guidewires under the same two model numbers, and Arrow marked guidewires (models PW-18080-HN and PW-16130-HN). The FDA's instruction is direct: stop use and distribution at once, review inventory, and quarantine affected lots.
The failure starts with water. The hydrophilic coating on the wire can swell inconsistently once saline hydrates it, and the swelling leads to clumping that occurs most often on the distal coil section. Clumped coating can flake, displace or detach, and detached particles can enter the bloodstream and cause a thromboembolic event, with the risk set by the size and quantity of particles released.
The wires are built from a nickel titanium alloy with a radiopaque metallic coil tip, and they're cleared for percutaneous entry of peripheral vessels with the Seldinger technique. They aren't indicated for the coronary or cerebral vasculature. The FDA notes that hydrophilic polymer embolization risks run highest in patients where embolized material could reach the heart, lungs, brain, kidneys, grafts or compromised peripheral circulation, and reported consequences in the alert range from silent deposits to stroke, organ failure and death.
That consequence list describes the hazard class, not this recall's record. As of November 5, 2026, Heraeus hadn't reported any serious injuries or deaths tied to this issue. The FDA says it's still reviewing the information and will update the alert as new findings arrive. Customers can reach Heraeus Medevio at [email protected] or 612-325-2003, and clinicians and patients can report problems through MedWatch.